AnaConDa
Anaesthetic Conserving Device

Cat. No. 26000

Description

AnaConDa is delivered in a blister package where each package contains:
• One AnaConDa, including closure caps
• One 50 ml Anaconda Syringe, including syringe closure
• Instructions for Use

The minimum order volume is a box of 6 units.

Operating Conditions

Operating Conditions
Disposable product:   To be replaced every 24 hours
Anaesthetic agents:   Isoflurane or sevoflurane
Efficiency:   Normally > 90% is re-circulated.
Depends on tidal volume and agent concentration.
Dosing:   See nomogram below
Tidal volume working range:   Minimum 350 ml
Resistance to gas flow @ 60 L/min:   2.5 cm H2O (250 Pa)
Moisture loss at 0.75 l x 12 breaths/min:   5 mg/l
Corresponding moisture output:   30 mg H2O/l
Moisture loss at 1.0 l x 10 breaths/min:   7 mg/l
Corresponding moisture output:   29 mg H2O/l
Bacterial filtration:   99.999%
Viral filtration:   99.98%
Dead space:   Approx. 100 ml
Connectors:   15F/22M-15M
Weight:   50g
Gas sampling port:   Female luer lock
Agent supply line:   2200 mm

Dosing

The nomogram provides a starting dose to be fine-tuned based on reading on the gas monitor. It provides the pump rate at different minute volumes to give 0,5% agent concentration. If another concentration is desired just multiply the pump rate with a ratio in relation to 0,5%.

Nomogram

Required syringe pump rate to reach 0,5% agent concentration at different minute ventilation volumes.

Downloads

Product leaflet [PDF, 196 KB]
(Also available in German in Downloads section)

Instructions for use [PDF, 148 KB]
(Also available in other languages in Downloads section)

Training Manual [PDF, 416 KB]
(Also available in Swedish in Downloads section)

Dosing Nomogram (in size A4) [PDF, 64 KB]

 

© 2010 Sedana Medical | About this site |